20
Calibration.
Reticle ruler.
Line measurement
Rectangle measurement.
Recovery to the default setting.
Set the color of the measure tools.
①
Dimension Calibration
Here the objective is 10X and the micrometer is 0.1mm.
1. Put a micrometer on the stage; select the objective (example here is 10X),
focus the microscope to make the micrometer display clearly on the LCD
screen.
2. Switch the command bar from “Image Effect Adjust” to “Measure”
Click
Measure