2460
Made in Taiwan
20170807 Version: 6.0
Notes:
Battery polarities must be correctly
installed. Otherwise, damage might be caused
to device.
4. Power Requirements: Two AAA 1.5V
Alkaline Batteries
Power consumption: less than 40mA
Low power indication:
Battery Life: Two AAA 1.5V, 600mAh
alkaline batteries could be continuously
operated as long as 30 hours.
5. Dimension:
Length: 61mm
Width: 32mm
Height: 37mm
Weight: 50g(including two AAA batteries)
6. Environment Requirements:
Operation Temperature: 5 ~ 40°C
Storage Temperature: -20 ~ 55°C
Ambient Humidity:
≤
80% no condensation
in operation
≤
93% no condensation
in storage
7. Declaration: EMC of this product comply
with IEC60601-1-1-2 standard.
8. Measurement Performance in Weak Filling
Condition: required the test equipment
(BIO-TEK INDEX Pulse Oximeter tester)
the pulse wave is available without failure
when the simulation pulse wave amplitude
is at 6%.
9. Interference Resistance Capacity aginst
Ambient Light: Device work normally when
mixed noise produced by BIO-TEK INDEX
Pulse Oximeter tester.
8. Product Accessories
1 One hang lace
2 Two batteries
3 One User manual
4 One drawsting bag
9. Battery Installations
1 Put the two AAA batteries into battery
cassette in correct polarities.
2 Push the battery cover horizontally along
the arrow shown as below:
Please put or remove batteries in right order,
or is likely to damage the device bracket.
Please remove the battery if the Oximeter will
not be used for long time
10. Hang Lace Installations
1. Thread thinner end of the hang lace
through the hanging hole
2. Thread thicker end of the lace through the
threaded end before pulling it tightly
11. Maintenance and Storage
12. Calibrating the pulse oximeter
14. Detailed descriptions of product
13. Declaration
1. Replace the batteries timely when low
voltage lamp is lighted.
2. Clean surface of the finger type Oximeter
before it is used in diagnosis for patients
3. Remove the batteries inside the battery
cassette if the Oximeter will not be
operated for a long time.
4. It is best to preserve the product in a
place where ambient temperatures range
15. Possible Problems and Resolutions
Problems
Possible
reason
Solution
1. Finger is not
plugged
correctly
2. Patient’s
Oxyhemoglobin
value is too
low to be
measured
1.Retry by
plugging the
finger
2. Try some more
times, If you can
make sure about
no problem
existing in the
product.
Please go to a
hospital timely
for exact
diagnosis
1. Finger might
not be plugged
deep enough
2. Finger is
trembling or
patient’s body
is in movement
status
1. Retry by
plugging the
finger
2. Try not to move
The Oximeter
can not be
powered on
1. Power of
batteries might
be inadequate
or not be there
at all
2. Batteries might
be installed
incorrectly
3. The Oximeter
might be
damaged
1. Please replace
batteries
2. Please reinstall
the batteries
3. Please contact
with local
customer service
centre
Indication
lamps are
suddenly off
1. The product
is automatically
powered off
when no signal
is detected
longer than 8
seconds
2. Power
quantity of the
batteries is
started being
inadequate
1. Normal
2. Replace the
batteries
“Error3” or
“Error4”
displayed on
screen
1 .Low power
2. Receiving
tube being
shielded or
damaged
together with
broken
connector.
3. Mechanical
misplace for
receive-emissi
on tube
4. Amp circuit
malfunction.
1. Change new
battery
2. Please
contact with
local customer
service center
3. Please
contact with
local customer
service center
4. Please
contact with
local customer
service center
17. Symbol Definitions
16. Guidance and manufacture’s declaration
- electronmagnetic emissions -
for all Equipmnet and Systems
5. It is recommended that the product should
be kept in a dry environment anytime.
A wet ambient might affect its lifetime
and even might damage the product.
6. Please follow the law of the local
government to deal with used battery.
1. The functional tester cannot be used to
assedd te accuracy of the oximeter.
2. Index 2 that made by Bioteck company is a
function tester. Set Tech to 1. R curve to 2,
then user can use this particular calibratio
curve to measure the oximeter.
3. The test methods used to establish the
SpO
2
accracy is clinical testing. The
oximeter used to measure the arterial
haemoglobin oxygen saturatin levels and
these levels are to be compared to the
levels determined from arterial blood
sampling with a CO-oximeter.
EMC of this product comply with IEC60601-1-
1-2 standsard.
The materials which user can come into
contact is no toxicity and no action on tissues,
comply with ISO10993-1 -5, -10.
1. Display Type: OLED display
2. SpO
2
:
Measurement range: 70%~100%
Accuracy: ±2% on the stage of 70%~100%
0%~69% no definition
3. Pulse Rate:
Measure range: 30~235 BPM
Accuracy: 30~99BPM, ±2 BPM ;
100~235BPM, ±2 %
SpO
2
or PR can
not be showed
normally
SpO
2
or PR is
showed unstably
“Error7”
“Error6’’
displayed on
screen
1. Low power
Err6 means the
screen is failure.
2. Emission
tube damaged.
3. Current
control circuit
malfunction.
1. Please
Please contact
local customer
service center.
change battery
2. Please
contact with
local customer
service center
3. Please
contact with
local customer
service center
functions
Guidance and manufacture’s declaration -
electromagnetic emission
The Pulse Oximeter is intended for use in the
electromagnetic environment specified below. The
customer of the user of the Pulse Oximeter should
assure that it is used in such and environment.
Emission test Compliance
Electromagnetic
environment-guldance
RF emissions
CISPR 11
Group 1
The Pulse Oximeter
uses RF energy only
for its internal function.
Therefor, its RF
emissionslare very low
and are not likely to
cause any interference
in nearby electronic
equipment.
RF emission
CISPR 11
Class B
The Pulse Oximeter is
suitable for use in all
establishments,
including domestic
establishments and
those directly
connected to the public
low-voltage power
supply network that
supplies buildings use
for domestic purposes.
Manufactured by:
Acare Technology Co., Ltd
6F.-3, No.24, Wuquan 2nd Rd.,Xinzhuang Dist.,
New Taipei City 242 Taiwan
TEL: +886-2-2298-8170
FAX: +886-2-2298-8560
http://www.acaretech.com
MEDIPRO
Villapark Business Park, Av Quitapesares 8
Building 8, Villaviciosa de Odon (Madrid) 28670
Spain
Symbol
Definition
The equipment type
is BF
Refer to user manual
before application
Hemoglobin saturation
Heart rate (BPM)
Low power indication
SN
Serial No. (Attached
inside of the unit )
Notes:
Please remove the battery when
product will not be in used for period of time.
No SpO
2
alarm
SpO
2
%
PR
bpm
i
from -20°C to 55°C and
≤
93% humidity.