2460

Made in Taiwan

20170807  Version: 6.0

 

 

Notes:

 

Battery  polarities  must  be  correctly 

installed. Otherwise, damage might be caused 
to device. 

4. Power Requirements: Two AAA 1.5V     

    Alkaline  Batteries

    Power consumption: less than 40mA

    Low power indication: 

    Battery Life: Two AAA 1.5V, 600mAh 

    alkaline  batteries could be continuously

    operated as long as 30 hours.

5. Dimension: 

    Length: 61mm

    Width: 32mm

    Height: 37mm

    Weight: 50g(including two AAA batteries)

6. Environment Requirements:

    Operation Temperature: 5 ~ 40°C 

    Storage Temperature: -20 ~ 55°C

    Ambient Humidity: 

80% no condensation

                                  in operation

                                  

93% no condensation

                                  in storage

7. Declaration: EMC of this product comply

    with IEC60601-1-1-2 standard. 

8. Measurement Performance in Weak Filling 

    Condition: required the test equipment

    (BIO-TEK INDEX Pulse Oximeter tester)

   the pulse  wave  is  available  without  failure

    when the simulation pulse wave amplitude 

    is at 6%.

9. Interference Resistance Capacity aginst

    Ambient Light: Device work normally when

    mixed noise produced by BIO-TEK INDEX

    Pulse Oximeter tester.

 

8. Product Accessories 

1 One hang lace 

2 Two batteries 

3 One User manual 

4 One drawsting bag 

9. Battery Installations 

1  Put  the  two  AAA  batteries  into  battery 

cassette in correct polarities. 

2 Push the battery cover horizontally along 

the arrow shown as below: 

 

 

 

 

Please put or remove batteries in right order, 

or is likely to damage the device bracket. 

Please remove the battery if the Oximeter will 

not be used for long time 

10. Hang Lace Installations 

1.  Thread  thinner  end  of  the  hang  lace 

through the hanging hole 

2.  Thread thicker end of the lace through the 

threaded end before pulling it tightly 

11. Maintenance and Storage 

12. Calibrating the pulse oximeter

14. Detailed descriptions of product 

13. Declaration 

1. Replace the batteries timely when low  

voltage lamp is lighted. 

2. Clean surface of the finger type Oximeter 

before it is used in diagnosis for patients 

3. Remove the batteries inside the battery 

cassette if the Oximeter will not be 

operated for a long time. 

4. It is best to preserve the product in a  

place where ambient temperatures range

 

15. Possible Problems and Resolutions 

 

Problems 

Possible 

reason 

Solution 

1. Finger is not 

plugged 

correctly 

2. Patient’s 

Oxyhemoglobin

value is too 

low to be 

measured 

1.Retry by 

plugging the 

finger 

2. Try some more 

times, If you can

make sure about

no problem 

existing in the 

product. 
Please go to a 

hospital timely 

for exact 

diagnosis 

1. Finger might 

not be plugged 

deep enough 

2. Finger is 

trembling or 

patient’s body 

is in movement 

status 

 

1. Retry by 

plugging the 

finger 

2. Try not to  move 

The Oximeter 

can not be 

powered on 

 

 

 

 

 

1. Power of 

batteries might 

be inadequate 

or not be there 

at all 

 

2. Batteries might

be installed

incorrectly 

 

3. The Oximeter 

might be

damaged 

 

1. Please replace 

batteries 

 

2. Please reinstall 

    the  batteries 

 

3. Please contact  

with local 

customer service  

centre 

 

Indication 

lamps are 

suddenly off 

1. The product 

is automatically 

powered off 

when no signal 

is detected 

longer than 8 

seconds 

2. Power 

quantity of the 

batteries is 

started being 

inadequate 

1. Normal 

 

2. Replace the 

batteries 

“Error3” or   

“Error4” 

displayed on 

screen 

1 .Low power 

2. Receiving 

tube being 

shielded or 

damaged 

together with 

broken 

connector. 

3. Mechanical 

misplace for    

receive-emissi

on tube   

 
4. Amp circuit 

malfunction. 

1. Change new 

battery   

 

2. Please 

contact with 

local customer 

service center 

3. Please 

contact with 

local customer 

service center 

 

4. Please 

contact with 

local customer 

service center 

 

 
 

 

 

 17. Symbol Definitions 

  

 

16. Guidance and manufacture’s declaration

- electronmagnetic emissions - 

for all Equipmnet and Systems 

 

5. It is recommended that the product should  

be kept  in  a  dry  environment  anytime. 

A  wet ambient  might  affect  its  lifetime  

and even might damage the product. 

6. Please  follow  the law of the local 

government to deal with used battery. 

1. The functional tester cannot be used to 

    assedd te accuracy of the oximeter.

2. Index 2 that made by Bioteck company is a

    function tester. Set Tech to 1. R curve to 2,

    then user can use this particular calibratio

    curve to measure the oximeter.

3. The test methods used to establish the 

    SpO

2

 accracy is clinical testing. The 

    oximeter used to measure the arterial 

    haemoglobin oxygen saturatin levels and 

    these levels are to be compared to the 

    levels determined from arterial blood  

    sampling with a CO-oximeter.

EMC of this product comply with IEC60601-1-

1-2 standsard.

The materials which user can come into 

contact is no toxicity and no action on tissues, 

comply with ISO10993-1 -5, -10.

1. Display Type: OLED display

2. SpO

2

 :

    Measurement range: 70%~100%

    Accuracy: ±2% on the stage of 70%~100%

                     0%~69% no definition

3. Pulse Rate: 

    Measure range: 30~235 BPM

    Accuracy: 30~99BPM, ±2 BPM ;     

                     100~235BPM, ±2 %

SpO

2

 or PR can 

not be showed 

normally

SpO

2

 or PR is 

showed unstably 

“Error7” 

“Error6’’

displayed on 

screen 

1. Low power 

Err6 means the 

screen is failure. 

 

2. Emission 

tube damaged. 

 

3. Current 

control circuit 

malfunction. 

1. Please 

Please contact 

local customer 

service center.

change battery 
2. Please 

contact with 

local customer 

service center 
3. Please 

contact with 

local customer 

service center 

functions 

Guidance and manufacture’s declaration - 

electromagnetic emission

The Pulse Oximeter is intended for use in the 

electromagnetic environment specified below. The 

customer of the user of the Pulse Oximeter should 

assure that it is used in such and environment.

Emission test    Compliance 

Electromagnetic 

environment-guldance

RF emissions

CISPR 11

Group 1

The Pulse Oximeter 

uses RF energy only 

for its internal function. 

Therefor, its RF 

emissionslare very low 

and are not likely to 

cause any interference 

in nearby electronic 

equipment.

RF emission

CISPR 11

Class B

The Pulse Oximeter is 

suitable for use in all 

establishments, 

including domestic 

establishments and 

those directly 

connected to the public 

low-voltage power 

supply network that 

supplies buildings use 

for domestic purposes.

 
 
   

Manufactured by: 

 

Acare Technology Co., Ltd 

6F.-3, No.24, Wuquan 2nd Rd.,Xinzhuang Dist., 

New Taipei City 242 Taiwan

TEL: +886-2-2298-8170

FAX: +886-2-2298-8560

http://www.acaretech.com    

    

MEDIPRO 

Villapark Business Park, Av Quitapesares 8

Building 8, Villaviciosa de Odon (Madrid) 28670

Spain

Symbol 

Definition 

The equipment type 

is BF 

Refer to user manual 

before application 

Hemoglobin saturation   

Heart rate (BPM) 

 

Low power indication 

SN

Serial No. (Attached 

inside of the unit ) 

Notes: 

Please remove the battery when 

product will not be in used for period of time.

No SpO

2

 alarm 

SpO

2

PR

bpm

i

from -20°C to 55°C and 

≤ 

93% humidity. 