background image

Cecomp

®

 2-Wire Loop Powered Digital Pressure Transmitters with Display Backup 

F16LSC, F16LSCN

Ranges and Resolution

See table below. Consult factory for special engineering units. 

Resolution is fixed as indicated in table.

Accuracy

Accuracy includes linearity, hysteresis, repeatability 

Standard accuracy:  ±0.25% of full scale ±1 least significant digit 

HA

 accuracy option: ±0.1% FS ±1 LSD, see ranges for availability 

Sensor hysteresis:  ±0.015% FS, included in accuracy 

Sensor repeatability: ±0.01% FS, included in accuracy

Display

4 readings per second nominal display update rate 

Up to 40 minute display operating time if loop power is lost 

4 digit LCD, 0.5" H and 5 character 0.25" H alphanumeric

Controls and Functions

SEL:  Press to momentarily display pressure if loop power is lost  

 

Press and hold to display pressure for up to 40 minutes 

 

depending on state of charge 

TEST:  Used with SEL button to set loop current to allow testing of 

 

system operation. May be pass code protected.

s t

  Up/down: When in setup used to set test values, enter pass 

 

code, units, scaling and calibration values

Calibration

User settable pass code required to enter calibration mode 

All pressure and absolute models: zero, midpoint, span 

All vacuum models: –span, –midpoint, zero 

Vacuum/pressure models: –span, zero, +midpoint, +span 

±15 psi models: –span, –midpoint, zero, +midpoint, +span

Loop Supply Voltage

Any DC supply/loop resistance that maintains 10-32 VDC at gauge 

terminals

Reverse polarity protected 

Approx. 2 hours required to charge backup capacitor

F16LSC

 

O

Display Pressure Even if Loop is Disconnected 

 

O

±0.25% Test Gauge Accuracy

 

O

316L Stainless Steel Wetted Parts

 

O

Scalable 4-20 mA Analog Output

 

O

Output Test Function

 

O

Selectable Engineering Units

Output Characteristics

4-20 mADC current output 

Passive transmitter, requires external loop power 

Output drive (compliance) determined by power source 

Output updated approximately 16 times per second 

12,000 counts over sensor range

Weight

9 ounces (approx.), shipping wt. 1 pound (approx.) 

Housing

F16LSC

: Extruded aluminum case, epoxy powder coated, ABS/ 

polycarbonate bezel (aluminum bezel optional), front and rear 

gaskets, polycarbonate label

F16LSCN

: ABS/polycarbonate NEMA 4X case, rear gasket, 

polycarbonate label
See other side for dimensions

Connection, Material, Media Compatibility

1/4" NPT male fitting, all wetted parts are 316L stainless steel

Overpressure, Burst, Vacuum

Overpressure: 2X pressure sensor range

Overpressure 3000 psig sensor: 5000 psig 

Overpressure 5000 psig sensor: 7500 psig

Burst pressure: 4X sensor rating, or 10,000 psi, whichever is less
Under-range display (non-vacuum sensors): 

–Err

Over-range display at 112.5% full scale: 

1– – –

 or 

I –.–.–.–

Vacuum service: 15 psia, ±15 psig, 15 psig, 30 psia, 100 psig,  

 

100 psia, 200 psig sensors

Environmental

Storage temperature: 

–40 to 203°F (–40 to 95°C)

Operating temperature: 

–4 to 185°F (–20 to 85°C)

Compensated temperature: 

32 to 158°F (0 to 70°C)

F16LSCN

How to Specify

Type

F16LSC

 range - options

Standard housing

F16LSCN

 range - options

NEMA 4X housing

 

psi = PSI 

  inHg = INHG 

 oz/in

2

 = ZIN 

 inH

2

O = INH2O 

 ftH

2

O = FTH2O 

  mmHg = MMHG 

 

torr = TORR 

 mmH

2

O = MMH2O 

 kg/cm

2

 = KGCM 

 g/cm

2

 = GCM 

 

kPa = KPA 

  MPa = MPA 

  mbar = MBAR 

 

bar = BAR 

 cmH

2

O = CMH2O 

 

atm = ATM

 

G = gauge reference pressure 

 

VAC = gauge reference vacuum 

 

A = absolute reference

Range

—see table at left

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

Made in USA

Quick Link 

cecomp.com/loop

NEMA 4X 

Housing

Options

—add to end of model number

HA

High accuracy, ±0.1% FS ±1 LS  

See table at left for availability

MC

Metal front cover instead of plastic, n/a NEMA 4X

PM

Panel mount, 4.1" x 4.1", n/a NEMA 4X

CC

Moisture resistant circuit board conformal coating

CD

Calibration data; 5 test points and date 

NC

NIST traceability documentation, 5 points and date

SCR14SS

Filter screen fitting keeps debris out of gauge 

sensor. Use for food vacuum packaging applica-

tions. 303 SS body, 100 micron 304 SS screen. 

© 07-19

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

Sensor Ranges and Engineering Units  

‡ -HA option not available

3 psig 

Res 15 psig vac 

Res

30 psia

Res

–15V100psig 

Res

300 psig

Res

3PSIG

.001

100KPAVAC

.1

2KGCMA

.001

–15V100PSIG

.1

300PSIG

.1

6INHGG

.001

0.1MPAVAC

.0001

2ATMA

.001

–30INHG/100PSIG

.1

610INHGG

.1

85INH2OG

.1

1BARVAC

.001

30 psig

Res

–30V200INHGG

.1

4800ZING

1

50ZING

.01

1KGCMVAC

.001

30PSIG

.01

–400V2770INH2OG

1

700FTH2O

.1

210GCMG

.1

1ATMVAC

.001

60INHGG

.01

–240V1600ZING

1

2000KPAG

1

150MMHGG

.1

15 psig

Res

850INH2OG

1

–760V5200MMHGG

1

2MPAG

.001

150TORRG

.1

15PSIG

.01

480ZING

.1

–760V5200TORRG

1

20BARG

.01

200MBARG

.1

30INHGG

.01

2100GCMG

1

–100V700KPAG

1

20KGCMG

.01

200CMH2OG

.1

400INH2OG

.1

1600MMHGG

1

–0.1V0.7MPAG

.001

20ATMG

.01

2000MMH2OG

1

240ZING

.1

1600TORRG

1

–1V7BARG

.01

500 psig

Res

7FTH2O

.001

1000GCMG 

1

2000MBARG

1

–1V7KGCMG

.01

500PSIG

.1

20KPAG

.01

760MMHGG

.1

2100CMH2OG

1

–1V7ATMG

.01

1020INHGG

1

5 psig 

Res

760TORRG

.1

70FTH2O

.01

100 psig

Res

1150FTH2O

1

5PSIG

.001

1000MBARG

1

200KPAG

.1

100PSIG

.1

3500KPAG

1

10INHGG

.01

1000CMH2OG

1

0.2MPAG

.0001

200INHGG

.1

3.5MPAG

.001

140INH2OG

.1

35FTH2O

.01

2BARG

.001

2770INH2OG

1

35BARG

.01

80ZING

.1

100KPAG

.1

2KGCMG

.001

1600ZING

1

35KGCMG

.01

350GCMG

.1

0.1MPAG

.0001

2ATMG

.001

7000GCMG

1

35ATMG

.01

260MMHGG

.1

1BARG

.001

60 psig

Res

5200MMHGG

1

1000 psig

Res

260TORRG

.1

1KGCMG

.001

60PSIG

.01

5200TORRG

1

1000PSIG

1

350MBARG

.1

1ATMG

.001

120INHGG

.1

7000MBARG

1

2040INHGG

1

350CMH2OG

.1

±15 psig 

Res

1660INH2OG

1

7000CMH2OG

1

2300FTH2O

1

3500MMH2OG

1

±15PSIG

.01

960ZING

1

230FTH2O

.1

7000KPAG

1

12FTH2O

.01

–30INHG/15PSIG

.01

4200GCMG

1

700KPAG

.1

7MPAG

.001

35KPAG

.01

±30INHGG

.01

3100MMHGG

1

0.7MPAG

.0001

70BARG

.01

15 psia

Res

±400INH2OG

1

3100TORRG

1

7BARG

.001

70KGCMG

.01

15PSIA

.01

±240ZING

.1

4100MBARG

1

7KGCMG

.001

70ATMG

.01

30INHGA

.01

±1000GCMG

1

4200CMH2OG

1

7ATMG

.001

2000 psig

Res

400INH2OA

.1

±760MMHGG

1

140FTH2O

.1

–15V200 psig 

Res

2000PSIG

1

240ZINA

.1

±760TORRG

1

400KPAG

.1

–15V200PSIG

.1

4070INHGG

1

1000GCMA

1

±1000MBARG

1

0.4MPAG

.0001

–30INHG/200PSIG

.1

4600FTH2O

1

760MMHGA

.1

±1000CMH2OG

1

4BARG

.001

–30V400INHGG

.1

14MPAG

.01

760TORRA

.1

±100KPAG

.1

4KGCMG

.001

–400V5500INH2OG

1

140BARG

.1

1000MBARA

1

±0.1MPAG

.0001

4ATMG

.001

–240V3200ZING

1

140KGCMG

.1

1000CMH2OA

1

±1BARG

.001

100 psia

Res

–100V1400KPAG

1

140ATMG

.1

100KPAA

.1

±1KGCMG

.001

100PSIA

.1

–0.1V1.4MPAG

.001

3000 psig

Res

0.1MPAA

.0001

±1ATMG

.001

200INHGA

.1

–1V14BARG

.01

3000PSIG

1

1BARA

.001

30 psia

Res

2770INH2OA

1

–1V14KGCMG

.01

6100INHGG

1

1KGCMA

.001

30PSIA

.01

1600ZINA

1

–1V 14ATMG

.01

6900FTH2O

1

1ATMA

.001

60INHGA

.01

7000GCMA

1

200 psig

Res

20MPAG

.01

15 psig vac 

Res

850INH2OA

1

5200MMHGA

1

200PSIG

.1

200BARG

.1

15PSIVAC

.01

480ZINA

.1

5200TORRA

1

400INHGG

.1

200KGCMG

.1

30INHGVAC

.01

2100GCMA

1

7000MBARA

1

5500INH2OG

1

200ATMG

.1

400INH2OVAC

.1

1600MMHGA

1

7000CMH2OA

1

3200ZING

1

5000 psig

Res

240ZINVAC

.1

1600TORRA

1

700KPAA

.1

480FTH2O

.1

5000PSIG

1

1000GCMVAC

1

2000MBARA

1

0.7MPAA

.0001

1400KPAG

1

35MPAG

.01

760MMHGVAC

.1

2100CMH2OA

1

7BARA

.001

1.4MPAG

.001

350BARG

.1

760TORRVAC

.1

200KPAA

.1

7KGCMA

.001

14BARG

.01

350KGCMG

.1

1000MBARVAC

1

0.2MPAA

.0001

7ATMA

.001

14KGCMG

.01

340ATMG

.1

1000CMH2OVAC

1

2BARA

.001

14ATMG

.01

Pb

Lead 

Free

Содержание Cecomp F16LSC

Страница 1: ...of model number HA High accuracy 0 1 FS 1 LS See table at left for availability MC Metal front cover instead of plastic n a NEMA 4X PM Panel mount 4 1 x 4 1 n a NEMA 4X CC Moisture resistant circuit board conformal coating CD Calibration data 5 test points and date NC NIST traceability documentation 5 points and date SCR14SS Filter screen fitting keeps debris out of gauge sensor Use for food vacuu...

Страница 2: ...15psi 4 mA 12 mA 20 mA Compound 30inHg 100psi 4 mA 5 5 mA 20 mA Compound 30inHg 200psi 4 mA 4 8 mA 20 mA Absolute reference 4 mA 4 mA 20 mA Bipolar 4 mA 12 mA 20 mA cecomp com 1220 American Way Libertyville IL 60048 800 942 0315 CECOMP Div of ABSOLUTE PROCESS INSTRUMENTS 1 65 2 88 NEMA 4X 3 0 2 0 Turn at hex fitting only Turn at hex fitting only 3 ft long 2 conductor 22 AWG cable 3 ft long 2 condu...

Страница 3: ...will be off The pressure will display for 15 seconds then the gauge will shut off SEL TEST t s SEL TEST t s on hold SEL TEST t s OFF SEL TEST t s i00 0 psig After initialization the gauge will indicate the applied pressure for 15 seconds or The gauge must be powered by a loop supply during configuration Release all buttons when the display indicates CFG The gauge firmware version is also dis playe...

Страница 4: ...l restarted manually or power is removed The display will then indicate the currently applied pressure in the engineering units selected in gauge configuration Button Operation Each time the s or t button is pressed and released quickly a small change is made to the digitized pressure signal It may take more than one of these small changes to result in a single digit change on the display To make ...

Отзывы: