SECTION 3 OPERATION
• In a case when incorrect operating parameters are applied to an
ozone generator the unit is switched off by the control system and
the LCD screen will indicate which of the parameters caused the
generator shutdown.
• In a safety shutdown mode the generator will be constantly moni-
toring operating parameters and the moment they are corrected the
generator will restart automatically, without any assistance.
• I.e. if there is a power spike or if there is a pressure drop caused
by temporary power failure and interruption of oxygen supply, the
moment the pressure is normal again or the voltage supplied is with-
in the operating range, the generator will restart and will produce
ozone again.
3B. System Startup
1. Make sure the Ozone Generator enclosure is securely attached to
an appropriate frame or wall Ozone Generators Stack.
2. Make sure all connections to the generator, power and gas are
made according to local codes and regulations.
3. Purge the generator with 90% or higher concentration oxygen and
assure a free flow of oxygen through the system and removal of con-
taminants. Conduct system leak test applying working pr/-
10% using oxygen only. Adjust gas pressure across the cell according
to specifications using downstream Control Valve (on flow meter)
and pressure gauge (supplied by the installer).
4. Plug the system into a specified receptacle.
5. Flip the Power Switch into the “ON” position.
6. Slide the red dot on the power control slider upwards to adjust to
desired power from 0-100%according to supplied “Ozone Generator
Performance Test” chart. Now the generator is at optimum
production mode and ozone should be flowing through the system.