Section 4
Instrument Installation
Abbott VP 2000™ Processor Service Manual
4-21
30-608308/R1—September 2008
5. Remove the CPU from its box and place it in the desired
location by following the guidelines as described
previously.
6. Plug the power cord into port A.
7. Connect the long RS232 data cable between serial data
port B on the back of the PC and data port C (“Computer
Interface”) on the right side of the Abbott VP 2000
Processor.
8. Connect the short RS232 data cable between data ports D
(“Stainer Interface”) and E on the back of the Abbott VP
2000 Processor.
9. Connect the PC power cord into a grounded outlet with
appropriate voltage.
10. Power ON the PC.
NOTE:
Any time you disconnect to reconnect the
RS232 data cable, be sure the two (2) units are
powered OFF.
DB9 Cable Attachment
1. Attach the DB9 male end of the included serial cable to the
serial port on the CPU.
2. Attach the DB9 female end of this serial cable to the port
on the Abbott VP 2000 Processor labeled “To Computer.”
UPS Assembly
1. Remove the UPS and attach to the nearest electrical outlet.
2. Attach the monitor and PC to the battery backup outlets
on the UPS.
3. Attach the Abbott VP 2000 Processor power cord to the
surge protection outlet on the UPS.
4. Follow the instructions for the UPS to ensure that it is
turned ON.
Содержание VP 2000 Processor
Страница 2: ...Introduction ii Abbott VP 2000 Procesor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 13: ...Abbott VP 2000 Processor Service Manual 1 1 30 608308 R1 September 2008 SECTION 1 ABOUT THIS MANUAL ...
Страница 14: ...About This Manual Section 1 1 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 22: ...About This Manual Section 1 1 10 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 23: ...Abbott VP 2000 Processor Service Manual 2 1 30 608308 R1 September 2008 SECTION 2 SAFETY Section 2 Safety ...
Страница 24: ...Safety Section 2 2 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 38: ...Technical Data Section 3 3 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 70: ...Installation Section 4 4 2 Abbott VP 2000 Proceessor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 97: ...Section 5 Purpose of This Section Abbott VP 2000 Processor Service Manual 5 7 30 608308 R1 September 2008 ...
Страница 144: ...Troubleshooting Section 6 6 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 146: ...Troubleshooting Section 6 6 4 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES TOC ...
Страница 270: ...Shutdown Storage and Shipping Section 8 8 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 280: ...Spare Parts and Accessories Section 9 9 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 288: ...Spare Parts and Accessories Section 9 9 10 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...