SOTOX A™ MOBILE ANALY ZER | USER GUIDE
6 | EN
1. INTRODUC TION
• Use the analyzer indoors when connected to AC power.
• Do not open the analyzer. The analyzer contains no user serviceable parts. Any unauthorised opening of the analyzer can cause hazards and
will invalidate warranty claims.
• The test cartridges are designed for single use only – do not attempt to re-insert a used test cartridge into the analyzer.
• Do not insert fingers or objects other than the test cartridge into the test cartridge slot. This may result in harm to the user or damage to the
analyzer.
• Do not use dirty or damaged test/QC cartridges as this may damage the analyzer or cause an invalid test result.
• Do not use the analyzer or any of its electrical accessories if these appear to be damaged. Contact Customer Service to arrange for
replacement parts.
• Do not use the analyzer near a magnetic source.
• Do not ingest the silica gel packs.
• Do not open power supply casing or printer.
• Allow the analyzer to reach ambient temperature before use. If moving the analyzer between storage and operating environments of
different temperatures or humidities, this may take at least 30 minutes.
If climatic differences are significant, allow more than 30 minutes for the analyzer to acclimatise.
1.9
Analyzer calibration
The SoToxa™ Mobile Analyzer is factory calibrated and does not require any further calibration by the user.
However, it is recommended to carry out a regular performance check using the quality control test cartridges provided with the SoToxa™
Mobile Test System (see section 8 QC Tests).
1.10 Maintenance and cleaning
If required, the exterior of the SoToxa™ Mobile Analyzer and its accessories can be wiped with a lint free cloth or a damp cloth with mild
detergent. Wipe away any spillages immediately. Do not use abrasive cloths or other material, and do not allow the unit to come into contact with
any potentially corrosive substances. If using a damp cloth, ensure the analyzer is turned off and disconnected from mains A/C power.
DO NOT try to clean inside the analyzer or open the analyzer for cleaning.
The analyzer does not contain any user serviceable components. If a fault is suspected or routine factory service is required, then please
contact Customer Service.
1.11 Servicing
The SoToxa™ Mobile Analyzer requires an annual service which must be carried out by an Abbott authorised and trained service engineer.
It is important to have your analyzer serviced annually to ensure it is performing correctly. This requires the analyzer and accessories to be
returned to the manufacturer or authorised servicing facility for a short period of time.
The date that the next service is due can be viewed in the Analyzer Summary screen, and the analyzer will automatically display the service
symbol in the status bar when service is due in less than one month. If the analyzer is not returned to the manufacturer for servicing within
this time, the symbol will turn from white to orange. Additionally, a warning symbol will be shown on the screen denoting that the service
is overdue and the user will be locked out from completing any further drugs tests.