10
2.
Select the Interrogate Monitors button.
3.
The programmer will look for advertising pulses from the Confirm Rx™ Insertable Cardiac
Monitor.
Merlin™ PSA Model EX3100 Setup
Connect the Merlin™ PSA Model EX3100 into the PSA port on the Merlin 2 PCS
For more information on the Merlin PSA Model EX3100, including connection of the patient
cables and patient-cable adapters, refer to the on-screen help on the Merlin 2 PCS and to the
Merlin PSA user's manual.
Figure 7. Merlin™ PSA Model EX3100
1.
Functional Status LEDs
2.
Pace and Sense LEDs
3.
Receptacle for patient cable or patient-cable adapter
4.
Connector to Merlin 2 PCS
RF Communication Setup
Some St. Jude Medical implantable devices can communicate with the Merlin™ 2 PCS via radio
frequency (RF), which allows for a greater distance between the device and the Merlin 2 PCS. To
establish RF communication, follow the instructions below.
1.
Connect the Merlin™ Antenna to the RF Telemetry port on the back of the Merlin 2 PCS
(see figure below).
NOTE: Connect the Merlin Antenna only to the RF Telemetry port on the back of the
programmer. Do not connect the Merlin Antenna to any other ports.