8 ID NOW Instrument User Manual
1.3.2 12V Power Supply
Rated input voltage:
100-240VAC
Rated input frequency:
50/60Hz
Rated input current:
1.2 Amps
Operating environment:
As per instrument
Storage environment:
As per instrument
Output voltage:
12V
Output current:
3.5 Amps
1.4 Unpacking
Unpack the ID NOW Instrument and place on a clean, flat, level and stable surface
within reach of an electrical outlet. The ID NOW Instrument has a carrying handle
located on the back of the instrument for transport purposes. Use the carrying
handle when removing the instrument from the box and when transporting short
distances. Place instrument in a location where the air vents located in the back of
the instrument will not be blocked. Avoid placing the instrument in direct sunlight.
Inspect for any obvious signs of damage. Report any damage immediately to Abbott.
Allow the instrument to equilibrate to room temperature for at least
30 minutes.
1.5 Instrument Calibration
The ID NOW Instrument is factory calibrated and does not require any further
calibration and verification.
However, if the instrument was transported or moved, a performance check
using ID NOW positive and negative controls is recommended to ensure proper
functionality.