Message
Cause
Action
Analyzer Error, Use
Electronic
Simulator
Handheld detects
problem from which it
is likely to recover.
Insert the Electronic Simulator. If
PASS, continue to use handheld.
Analyzer Error,
See Manual
Handheld detects
problem from which it
may not recover.
Insert Electronic Simulator. If PASS,
insert a cartridge with sample
or control. If the code does not
reappear, continue to use handheld.
No display
Disposable batteries
dead or rechargeable
battery fully
discharged. Keypad
not responding. Start
switch broken.
Change disposable batteries or
recharge battery. If still no display,
call Support Services.
Cartridge Locked
does not disappear
after test cycle
completed
Dead battery(s).
Mechanical
problem.
Wait until handheld turns off. Turn
handheld on. If resets, remove
cartridge. If not, change or recharge
battery(s) and turn handheld on.