Replacing the Strip Port Module
The strip port module is designed to minimise the potential for liquid or debris to enter the meter through the strip port. Should
blood, liquid or other contaminant enter into the strip port, the strip port module must be replaced. Make sure the area around
the strip port is completely dry prior to installing the replacement strip port module.
Removal
1. Turn the FreeStyle Precision Pro meter off by pressing the On/Off
button
on the keypad.
2. Place the meter, display screen down, on a flat surface. Place one hand on the
handle of the meter to secure it. With the other hand use a small cross head
(Phillips) screwdriver (not included) to remove the single screw on the back of the
strip port module. (Fig. 1)
3. Once the screw is removed, place it aside. Use your hand to carefully slide the
strip port module away from the meter. (Fig. 2)
4. Remove the strip port module entirely. Discard it and the screw in accordance
with your facility’s biohazard disposal policy.
Figure 1
Figure 2
IMPORTANT:
Do NOT replace batteries and strip port module at the same time. Ensure batteries are
installed before replacing strip port module. Rechargeable batteries must be at room
temperature before they are placed in the battery compartment. The docking station
does not charge the batteries.
Chapter 12: Maintenance
Содержание FreeStyle Precision Pro
Страница 1: ...Operator s Manual for Healthcare Professionals...
Страница 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Страница 98: ...7 Linearity Test...
Страница 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...
Страница 172: ...12 Maintenance Cleaning the Exterior Surface Replacing the Strip Port Module Replacing the Batteries...
Страница 179: ...13 Specifications FreeStyle Precision Pro Meter Symbols Barcode Types...