![Abbott FreeStyle Libre Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/abbott/freestyle-libre/freestyle-libre_user-manual_3052262057.webp)
Control Solution Testing
You should do a control solution test when you are not sure of your test
strip results and want to check that your Reader and test strips are working
properly . You can do a control solution test with a blood glucose or ketone
test strip .
IMPORTANT:
•
Control solution results should fall within the control
solution range printed on the test strip instructions for use .
•
Do NOT use control solution past the expiry date . Discard
control solution three months after opening . Refer to control
solution instructions for use .
•
The control solution range is a target range for control
solution only, not for your blood glucose or ketone results .
•
The control solution test does not reflect your blood glucose
or ketone levels .
•
Use only MediSense glucose and ketone control solution .
•
Check that the LOT number printed on the test strip foil
packet and instructions for use match .
•
Replace the cap securely on the bottle immediately after use .
•
Do NOT add water or other liquid to the control solution .
•
Contact Customer Service for information on how to obtain
control solution .