Circulatory Support System Operation Manual (US)
Page 66 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
This page contains information about the software installed on the Console.
4) CHARACTER SET (FOR SERVICE PURPOSE ONLY)
This page displays all the possible characters that the Console display can produce.
To exit from the Console BIOS, turn the System OFF using the main power switch on the
rear panel of the Console.
7.9 Setting the Console Max Flow Alert
Main Menu
Minimum Flow Alert
Maximum Flow Alert
Pressure Menu
Stopwatch*
Extended Menu
Set RPM
Exit
Menu
Increase
Decrease
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Menu
Menu
Menu
Menu
Menu
Menu
Increase
Decrease
Enter
Enter
[Start|Stop|Reset]
Stop Watch 2
[Start|Stop|Reset]
Stop Watch 1
* Monitor only
Figure 36: Monitor MENU Structure
Setting Maximum Flow Alert
To change the
MAXIMUM FLOW ALERT
setting, depress the
MENU
keypad until the
MAX FLOW ALERT
option is displayed. Use the
DECREASE
or
INCREASE
arrow keys
to decrease or increase the alert setting. The default
MAX FLOW ALERT
value is set to
8.0 LPM at startup. The
MAX FLOW ALERT
threshold setting is continuously displayed in
the upper part of the display.
If the System detects flow in excess of 10 LPM, the digital flow display will not display a
numerical flow, but will display upwards pointing arrows: “^^.^^ LPM”