CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 114 of 126
© 2019 Abbott– Document No PL-0047, Rev 11 (December 2019)
12 APPENDICES
12.1 Appendix I – Console Alarms and Alerts
Table 16: Console Alarms & Alerts
ID
Alarm/
Alert
Text Message
System Status & Operator Response
S1
Alarm
POWER ON TEST FAIL
The Pump will not start.
An audible alarm will sound, which cannot be muted.
Switch the Console OFF and ON again. If the alarm re-appears,
switch to the backup Console, Motor and Flow Probe, record
the alarm message and contact your local Abbott Medical
representative.
S2
Alarm
SYSTEM FAULT
(Run-Time System Failure)
The Pump will stop.
An audible alarm will sound, which cannot be muted.
Clamp the return tubing and switch to the backup Console,
Motor and Flow Probe according the procedure described in
Section 10.1. Resume support. Record the alarm message and
contact your local Abbott Medical representative.
M1
Alarm
MOTOR STOPPED
The Pump will stop.
An audible alarm will sound, which can be muted for 60
seconds.
Clamp the return tubing and switch to the backup Console,
Motor and Flow Probe according the procedure described in
Section 10.1. Resume support. Record the alarm message and
contact your local Abbott Medical representative.
M2
Alarm
MOTOR DISCONNECTED
The Pump will stop.
An audible alarm will sound, which can be muted for 60
seconds.
During setup of the System:
Press the alarm acknowledge button and check that the Motor
connector is fully inserted into the back of the Console.
During support:
Press the alarm acknowledge button and check that the
connector of the Motor is fully inserted into the back of the
Console. Resume support. If the visual alarm message does not
disappear, clamp the return tubing and switch to the backup
Console, Motor and Flow Probe according the procedure
described in Section 10.1. Resume support. Record the alarm
message and contact your local Abbott Medical representative.
M3
Alarm
PUMP NOT INSERTED
System will not start.
An audible alarm will sound, which can be muted for 60
seconds.
Press the alarm acknowledge button. Insert or re-insert the
Pump and secure it with the locking screw.
If the alarm repeats, switch to backup Console, Motor and
Flow Probe according the procedure described in Section 10.1.