CELL-DYN
®
3200 System Operator’s Manual
3-5
9140181H—October 2001
Section 3
Principles of Operation
Fragile WBCs and Resistant RBCs
When running samples in the normal Patient mode, the operator may suspect the
presence of fragile WBCs when the FWBC flag is displayed or may suspect the
presence of resistant RBCs when the RRBC and NRBC flags are displayed.
In the case of samples containing fragile WBCs or resistant RBCs, an alternate
method is used to measure white blood cells. The results of this method are referred
to as the Nuclear Optical Count (NOC). The NOC measurement is derived from the
HGB dilution as described below. Refer to
Nuclear Optical
Count
and
Resistant
RBC
later in this section for additional information.
There are two different key labels on the
SPECIMEN TYPE
screen, one for fragile
WBCs and one for resistant RBCs. Refer to
Run Menu Soft Keys,
in
Specimen Types available.
When the Fragile WBC specimen type is selected, both NOC and WOC are
reported in the Data Log. The NOC value is reported as WBC on the
RUN
screen
and Laboratory Worksheet.
When the Resistant RBC specimen type is selected, both NOC and WOC are
reported in the Data Log. Either NOC or WOC is
reported as WBC (based on
algorithmic decision-making) on the RUN screen and Laboratory Worksheet.
NOTE:
When QC specimen type is selected, both NOC and WOC are reported in
the Data Log. Either NOC or WOC is reported as WBC (based on
algorithmic decision-making) on the RUN screen.
The analysis for Fragile WBC and Resistant RBC is performed as follows:
1. After the HGB sample is measured (refer to
Hemoglobin Analysis
earlier in
this section), the Sample Transfer Pump transfers the diluted solution from
the HGB Flow Cell to the Optical Flow Cell Sample Feed Nozzle.
2. Diluent/Sheath reagent, under constant pressure in the Sheath Reservoir, is
directed into the Optical Flow Cell.
3. Sequentially, the Sample Metering Syringe injects 140 µL of the HGB
dilution into the flow cell.
4. The higher speed of the sheath which surrounds the HGB dilution, and the
special geometry of the flow cell, focus the HGB dilution stream so that
individual cells can be counted.
5. A laser beam is focused on the flow cell. As the sample stream intersects the
laser beam, the light scattered by the cells is measured by the 0 degree
detector. The nuclei of the lysed cells are counted as the NOC result.
6. The WBC Analysis in Fragile WBC and QC specimen type occurs as
described in
WBC Analysis
earlier in this section.
7. The WBC Analysis in Resistant RBC mode occurs as described in
WBC
Analysis
above except that the diluted WBC segment is lysed in the WBC
Mixing Chamber for an additional 15 seconds.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...