3-4
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Principles of Operation
Sample Analysis Cycle Overview
Section 3
Hemoglobin Analysis
1. The Diluent/Sheath Syringe injects 1.7 mL of diluent through the Shear
Valve where the 12 µL HGB volume is transferred to the HGB Flow Cell.
2. The HGB Lyse Syringe dispenses 0.9 mL of HGB Lyse into the line after the
diluent has transferred the HGB volume to the HGB Flow Cell. The entry
point for the HGB Lyse is between the Shear Valve and the HGB Flow Cell.
3. The segment, lyse, and diluent are routed to the HGB Flow Cell where the
dilution is swirl mixed. The final dilution is 1:218.
4. A low-energy LED attached to the HGB Flow Cell measures the absorbance
of light at 555 nm. The absorbance is proportional to the HGB concentration
of the sample.
WBC Analysis
WBCs are analyzed optically as follows:
1. The WBC Lyse Syringe dispenses 0.973 mL of WBC Lyse reagent through
the shear valve where the 20 µL WBC volume is transferred to the WBC
Mixing Chamber.
2. The segment and reagent are then routed to the WBC Mixing Chamber where
the dilution is swirl mixed. The final dilution is 1:50. The diluted sample
remains in the mixing chamber for 14 seconds for the lysing of the red blood
cells.
3. The Sample Transfer Pump transfers the WBC dilution from the WBC
Mixing Chamber to the Optical Flow Cell Sample Feed Nozzle.
4. Diluent/Sheath reagent, under constant pressure in the Sheath Reservoir, is
directed into the Optical Flow Cell.
5. Sequentially, the Sample Metering Syringe injects 46.5 µL of the WBC
dilution into the flow cell at a pressure (and speed) lower than that of the
diluent/sheath reagent.
6. The higher speed of the sheath, which surrounds the WBC dilution, and the
special geometry of the flow cell combine to focus the WBC dilution stream
so that individual cells can be counted.
7. A laser beam is focused on the flow cell. As the sample stream intersects the
laser beam, the light scattered by the cells is measured at four different
detectors located in the forward (0° and 10°) and side (90° and 90°D) angles.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...