Quality Control Guide
The CELL-DYN 3200 System offers several quality control options to monitor and
validate instrument performance while running the Reticulocyte Package. The
options are:
•
6 QC Files
Statistical and graphical analysis of the data in each
file to calculate the mean, standard deviation, and
coefficient of variation
•
Westgard
®
Rules
A multi-rule system applied to the data in each of the
QC files
Each of these options is discussed in detail in
All QC data should be reviewed according to your laboratory’s protocol.
Control Material
Abbott Diagnostics recommends using the CELL-DYN Retic-RITE
™
control
material for performing quality control checks on the CELL-DYN 3200 System.
These controls should be run:
• After daily start-up procedures are completed.
• After a reagent lot number change.
• After a service call or component replacement.
• After calibrating the Standard Hematology Mode.
• In accordance with the laboratory’s quality control protocol.
• According to regulatory requirements.
NOTE:
Data invalidating alerts, such as Fragile RBC or ERL, are not valid
when running commercial controls.
Mixing and Handling
CAUTION:
When using the reticulocyte reagent, avoid contact with skin
and clothing. This reagent contains New Methylene Blue, which will stain
skin, clothing, and many other surfaces.
Reagent
• Use Reticulocyte Reagent prepared only by Abbott Laboratories. Verify the
expiration date.
• Store the Reticulocyte Reagent in the dark at room temperature.
• Use one Reticulocyte Reagent tube for each CELL-DYN Retic-RITE™
control or patient specimen.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...