CELL-DYN
®
3200 System Operator’s Manual
13-21
9140181H—October 2001
Section 13
Sample Loader
Operating Procedures
1. Press the
[START LOADER]
key in the
RUN
menu. Based on which condition
exists (refer to the
Sample Loader Description
,
Soft Keys
subsection), either
the Sample Loader will be activated or a submenu will be displayed.
2. If the submenu is displayed, press either
[RESUME LOADER]
or
[RESET
LOADER]
(refer to the
Sample Loader Description
,
Soft Keys
subsection for
a discussion of when to use each key).
3. Racks are moved in sequential order from the load side, through the Mixing
and Aspiration Stations to the unload side (refer to Figure 13.8).
4. For sample mixing and aspiration to occur, the sensors must detect a tube in
the rack. The Sample Loader automatically stops after the last rack is
processed and no more racks are detected on the load side.
5. When the tube is at the Vent/Aspiration Station, the Spinner positions the
tube vertically in the rack, spins the tube for bar code reading, and holds the
tube while it is vented and the sample aspirated. Proper positioning is
necessary for the tube stopper to be pierced in the center.
6. The Tower
Cover on the CELL-DYN 3200SL model fits over the Mixing
Station and Bar Code Reader on the loader. This cover prevents aerosol
contamination and exposure to the needle while the Sample Loader is
processing specimens.
NOTE:
The Sample Loader has an interlock switch that prevents operation
when the Tower Cover is not in place. If the Tower Cover is
removed during Sample Loader operation, processing is
immediately halted. For an orderly stop, the operator must press the
[STOP LOADER]
key and wait for the Sample Loader and Analyzer
to stop processing before lifting or removing the cover. Lifting the
cover while the Sample Loader is operating causes an immediate
Emergency Stop condition. The entire system must be re-initialized.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...