![Abbott CELL-DYN 3200 Скачать руководство пользователя страница 522](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985522.webp)
10-88
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
>>>> In Place of Results for WBC, RBC, HGB or PLT
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. Data exceeds linear range for that
parameter.
1. Dilute the sample with diluent/
sheath according to your laboratory
procedures and rerun the sample.
Multiply the results by the dilution
factor.
2. Improper reagent delivery or sample
aspiration.
2. Inspect syringes for proper
operation and check reagent Inlet
Lines for crimping.
High Rate of False Flagging
High number of false flags displayed on the
RUN
screen.
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. Dirty Diluent/Sheath Filter.
1. Replace Diluent/Sheath Filter.
2. Contaminated WBC Lyse.
2. Replace WBC Lyse reagent. Clean
WBC lyse syringe.
3. Dirty optical flow cell.
3. In S
PECIAL PROTOCOLS
menu,
follow procedure for emptying and
refilling optical flow cell.
4. Debris is present in the system.
4. In the second
SPECIAL PROTOCOL
menu screen, perform an
[
AUTO
CLEAN
]
and/or
[
EXTENDED
AUTOCLEAN
]
cycle to clean the
system.
5. Optical gains have shifted.
5. Contact your Abbott Hematology
Customer Support Center.
Table 10.3 Sample-Related Fault Conditions
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...