Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Loader Fault 45: unload zone full detector malfunction
/BL
The system failed to retract mechanical arms in unload zone.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A mechanical problem has
prevented the rack arms in the
unload zone from retracting.
1. Check for an obstruction that is
preventing the rack arms in the
unload zone from retracting and, if
one is found, remove it. Then, reset
the racks and Sample Loader
operation by pressing the following
keys in order:
[START LOADER]
,
[RESUME LOADER]
.
2. A failure of the sensor or sample
loader mechanics has occurred.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Loader Fault 46: unload zone nearly full detector malfunction
/BL
The system is detecting an unload zone nearly full condition (4 racks in unload
zone) even though there are fewer than 4 racks in the unload zone.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A mechanical problem has
prevented the rack arms in the
unload zone from retracting.
1. Check for an obstruction that is
preventing the rack arms in the
unload zone from retracting and, if
one is found, remove it. Then, reset
the racks and Sample Loader
operation by pressing the following
keys in order:
[START LOADER]
,
[RESUME LOADER]
.
2. A failure of the sensor or sample
loader mechanics has occurred.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...