Download flow sequence timeout
/BL
The allotted time for flow sequence downloading was exceeded.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. The flow sequence did not compile,
possibly due to interruption by a
fatal fault, an improperly seated
Tower Cover or door, or a fluidics
problem.
1. Reboot the System.
NOTE:
If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Command transmission timeout
The Data Module did not receive acknowledgment of a command sent to the
Analyzer within the allotted time.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A failure in the Analyzer hardware,
possibly due to an unplugged or
damaged HSSL cable, has occurred.
1. Check the HSSL cable connections.
Then, initialize the Analyzer by
pressing the
[INITIALIZATION]
key
on the second
DIAGNOSTICS MENU
screen. If the message appears
again, reboot the System.
NOTE:
For the location of the
HSSL ports, refer to
NOTE:
If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. A software error has occurred.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...