Section 10
Troubleshooting and Diagnostics
Data acquisition overlap
/BL
The error may occur when several tasks are requested in rapid sequence. For
example, print data log, transmit result, sample processing, print ticket, print
report, etc.
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The timing of communication
between the Analyzer and the
System Software is incorrect.
1. Record what happened when the
message was displayed.
2. Initialize the Analyzer by pressing
the
[INITIALIZATION]
key on the
second
DIAGNOSTICS
MENU
screen.
Continue processing samples.
3. If the fault recurs, contact your
Abbott Hematology Customer
Support Center.
List mode data phase error
/BL
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The order of the data received for
display on the screen was incorrect.
1. Record what happened when the
message was displayed.
2. Initialize the Analyzer by pressing
the
[INITIALIZATION]
key on the
second
DIAGNOSTICS
MENU
screen.
Continue processing samples.
3. If the fault recurs, contact your
Abbott Hematology Customer
Support Center.
Message acknowledgment time out
/BL
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. Communication between the
Analyzer and the Data Module did
not occur when expected.
1. Record what happened when the
message was displayed.
2. Initialize the Analyzer by pressing
the
[INITIALIZATION]
key on the
second
DIAGNOSTIC MENU
screen.
Continue processing samples.
If the fault
recurs, contact your
Abbott Hematology Customer
Support Center.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...