CELL-DYN
®
3200 System Operator’s Manual
10-43
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Diluent/Sheath, HGB Lyse, or WBC Lyse empty/
SB and BL
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. Container is empty.
1. Install a new container of reagent.
Press
[CLEAR FAULT]
to resume
operation.
NOTE:
Do not pour any remaining reagent into the new container.
2. Reagent inlet tubing is crimped or
obstructed.
2. Inspect the whole length of the inlet
tubing to ensure it is not crimped
and/or remove any obstruction.
3. Reagent line is not on the bottom of
the container.
3. Ensure that the line is properly
inserted in the container and the
sinker is on the bottom of the
container.
4. An incorrect reagent or a
nonconductive liquid is connected to
the inlet tube.
4. Check the label on the reagent
container to be sure the correct
reagent is installed. Trace the line to
the inlet connector and ensure that it
is connected to the correct one.
Check the connection to be sure it is
secure and then press
[CLEAR
FAULT]
.
5. Appropriate syringe not mounted
properly.
5. Check proper mounting of syringes.
6. The Luer fitting connection on the
top of the appropriate syringe is
loose.
6. Verify the Luer connection on the
top of syringes is secure.
7. WBC Lyse or HGB Lyse only:
tubing in solenoid valve is pinched
or not fully inserted.
7a.Check tubing in N/C valves 23
(WBC Lyse) or 28 (HGB Lyse) is
not pinched or obstructed; replace
as necessary.
7b.Verify tubings are fully inserted in
solenoids 23, 25 (WBC Lyse) and
28, 24 (HGB Lyse).
8. Circuitry malfunction.
8. Contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...