9-60
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Nonscheduled Maintenance Procedures
Section 9
Figure 9.23 Y Valve Cleaning
9. Disconnect this tubing at the Shear Valve and place the end in the beaker to
catch liquid.
10. Slowly inject the cleaning solution into the Y Valve. The liquid should pass
through the Y Valve carrying any obstructing particles with it.
11. Repeat the flushing process if necessary. When the solution flows freely
through the valve, the flushing process is completed.
12. Replace the syringe containing cleaning solution with the syringe containing
the deionized water.
NOTE:
Use paper towels to absorb any liquid that is discharged by the
tubing.
13. Slowly inject deionized water into the Y Valve to remove any bleach residue.
14. When finished, remove the syringe containing deionized water.
15. Reconnect the tubing from the Y Valve to the top of the Open Sample Probe.
16. Reconnect the tubing between the Y Valve and the Shear Valve.
17. Remove the beaker.
18. Close the Right Front Cover and press into place. Reinstall Tower Cover.
19. In
[SPECIAL PROTOCOLS]
press
[ENABLE ANALYZER]
followed by
[RETURN]
and
[MAIN]
to return to the
MAIN MENU
.
20. Press
[RUN]
followed by
[SPECIMEN TYPE]
and
[BACKGROUND]
.
21. Run at least three background counts. Check that the background counts are
acceptable before running controls or patient samples.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...