9-52
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Nonscheduled Maintenance Procedures
Section 9
10. Reattach the Tower Cover (Tower Door on the CS model must be open) and
press into place.
11. Press
[ENABLE ANALYZER]
followed by
[RETURN]
and
[MAIN]
to return to the
MAIN MENU
.
12. Run at least three background counts to rinse the system. Check that the
background counts are acceptable before running controls or patient samples.
If the counts are unacceptable, troubleshoot accordingly (refer to
Troubleshooting and Diagnostics
13. Record this maintenance in your maintenance log.
Sample Transfer Pump Tubing Replacement
The tubing under the Sample Transfer Peristaltic Pump needs to be replaced on a
regular basis to ensure proper fluid movement through the instrument. This tubing
should be replaced at least every 2000 cycles. However, the frequency of
replacement depends on instrument use in each laboratory. The Sample Transfer
Pump is shown in Figure 9.6.
Materials Required
1. Sample Transfer Peristaltic Pump Tubing
2. Gloves, lab coat, and safety glasses
Procedure
1. In the MAIN MENU, press
[SPECIAL PROTOCOLS]
followed by
[DIS/ENAB
ANALYZER]
and
[DISABLE ANALYZER]
2. Gently pull on the hand hold of the Left Front Cover and swing the cover
open.
3. Locate the Sample Transfer Peristaltic Pump on the left side of the panel.
(Refer to Figure 9.4.)
4. Push the Pump Shoe away from the pump wheel and slip the tubing out.
5. Remove the tubing from the metal brackets that hold it on each side of the
rollers. (Refer to Figure 9.6.)
6. Disconnect the tubing at the plastic connector.
7. Connect the new tubing to the plastic connector.
8. Place the collars on the ends of the tubing into the metal guides. Hold the
Pump Shoe open and push the tubing underneath the pump rollers.
NOTE:
Position the tubing in the center of the rollers.
9. Close the Left Front Cover.
10. Press
[ENABLE ANALYZER]
to bring the instrument to the READY state.
11. Press
[RETURN]
followed by
[MAIN]
and
[RUN]
to display the
RUN
screen.
12. Run at least three background counts to rinse the system. Check that the
background counts are acceptable before running controls or patient samples.
If the counts are unacceptable, troubleshoot accordingly (refer to
Troubleshooting and Diagnostics
13. Record this maintenance in your maintenance log.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...