CELL-DYN
®
3200 System Operator’s Manual
9-47
9140181H—October 2001
Section 9
Service and Maintenance
5. Locate the tubing just above solenoid #24 that attaches to the HGB Flow Cell.
(Refer to the previous figure.) Attach a hemostat to the tubing between the
fitting and the HGB Flow Cell.
6. Disconnect the tubing from the fitting and insert the syringe. Remove the
hemostat. Dispense the cleaning solution into the flow cell.
7. Allow the bleach solution to remain in the flow cell for 5 minutes.
8. When the time has elapsed, aspirate the solution back into the syringe and
reattach the hemostat.
9. Reconnect the tubing and remove the hemostat.
10. Reinsert the plastic drip pan by aligning the grooves on the pan with the
upright support arms and sliding the pan down until it rests on the support.
11. Reattach the Tower Cover (Tower Door on the CS model must be open) and
press into place. Close Right Front Cover.
12. Press
[ENABLE ANALYZER]
followed by
[RETURN]
and
[MAIN]
and
[RUN]
.
13. Run at least three background counts to rinse the system. Check that the
background counts are acceptable before running controls or patient samples.
If the counts are unacceptable, troubleshoot accordingly (refer to
Troubleshooting and Diagnostics
14. Record this maintenance in your maintenance log.
Bar Code Reader Window Cleaning
The cleaning procedure is similar for the SL and CS models even though the
location of the Bar Code Reader is different.
Materials Required
1. Gloves, lab coat, and safety glasses
2. Applicator swabs (non-sterile)
3. Microscope lens tissues
4. Isopropyl alcohol, lens cleaner, or deionized water
Procedure
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
followed by
>
DIS/ENAB
ANALYZER]
and
[DISABLE ANALYZER]
.
2. Take three applicator swabs and wrap a lens tissue around each swab.
3. Moisten one of the wrapped swabs in isopropyl alcohol, lens cleaner, or
deionized water.
4. Locate the Bar Code Reader Window.
5. Wipe the window with the moistened swab. Using one of the dry wrapped
swabs, wipe the window dry. Use the second dry swab if necessary.
6. Visually inspect the window to ensure that blood, debris, and smudges have
been removed.
7. Press the
[ENABLE ANALYZER]
key to bring the system to the READY state.
Press
[RETURN]
followed by
[MAIN]
and
[RUN]
to return to the
RUN
screen.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...