![Abbott CELL-DYN 3200 Скачать руководство пользователя страница 217](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985217.webp)
CELL-DYN
®
3200 System Operator’s Manual
5-65
9140181H—October 2001
Section 5
Operating Instructions
Run Menu
The
RUN
menu and its sub-menus are described below, including soft keys,
demographics area, Work List, and bulletin line.
To display the
RUN
menu, press
[RUN]
in the
MAIN MENU
. If the instrument has not
been primed, the priming cycle occurs before the
RUN
menu is displayed.
RUN Screen Format
Upper Left Corner (Run Screen Demographics)
The Demographics Section of the
RUN
screen differs depending on the specimen
type selected. These differences are discussed below.
Patient
The Upper Left Corner of the
RUN
screen for patient samples displays the
following data entry fields:
NEXT ID—
used to enter the ID number for the next sample to be run. (Up to 12
characters may be entered.)
NAME—
used to enter the patient’s name. (Up to 28 characters may be entered.)
PAT ID—
used to enter patient identification information, such as a Social Security
number. (Up to 16 characters may be entered.)
SEX—
used to enter the sex of the patient.
DOB—
used to enter the date of birth of the patient. (4 digits must be entered for the
Year.)
COL—
used to enter the collection date and time of the specimen.(4 digits must be
entered for the Year.)
DR—
used to enter the name of the patient’s physician. (Up to 22 characters may be
entered.)
CMT—
used to enter operator comments. (Up to 16 characters may be entered.)
PARAM SET/LIMITS SET—
displays the number of the Parameter Set (1–4) and
Limit Set (1-6) that will be applied to the sample results
X-B Status—
If the XB-RBC and/or the XB-WBC functions are enabled, the XB
file status for each is displayed under the Status Box.
NOTE:
The Parameter and Limits sets applied to the sample may be changed after
the sample has been run. Refer to the description of the Data Log
[EDIT
SPECIMEN]
key given in the
Using the Data Log
later in this section.
An example of the
RUN
screen for patient specimens is shown in Figure 5.27.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...