Start-Up Procedure
To startup the instrument, do the following:
1. If the instrument is OFF, first turn the power ON. When
INITIALIZED
is
displayed in the
MAIN MENU
Status Box, press
[RUN]
to prime the instrument
and bring it to the READY state.
2. If the instrument is already ON and
STANDBY
or
INITIALIZED
is
displayed in the
MAIN MENU
Status Box, press
[RUN]
to prime the instrument
and bring it to the READY state.
3. Perform the daily Quality Control checks as directed within this section.
Daily Start-Up Procedure
The daily start-up routine consists of the following procedures:
1. Check the reagent levels and replace reagent containers as necessary.
2. Check printer paper; add more paper as necessary.
3. Check tubing in the Normally Closed Valves and Sample Transfer Peristaltic
Pump for breaks, crimps, or other obstructions.
4. Run background counts until acceptable results are obtained for all
background parameters (WBC, RBC, HGB, and PLT). Acceptable results are
specified in
Performance Characteristics and Specifications
.
5. Quality control procedures (which confirm calibration) should be performed
on a daily basis according to the laboratory’s protocol. Commercial control
materials should be properly warmed and mixed according to the
manufacturer’s recommendations. Patient controls should be handled
according to the laboratory’s protocol.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...