![Abbott CELL-DYN 3200 Скачать руководство пользователя страница 177](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985177.webp)
CELL-DYN
®
3200 System Operator’s Manual
5-25
9140181H—October 2001
Section 5
Operating Instructions
Figure 5.9 QC Range Entry Screen
Range Entry
1. Select a file from the
QC SET UP
screen by using the Arrow keys on the
keyboard to move the cursor into the desired file.
2. Press
[QC LIMITS]
to display the
QC RANGE ENTRY
screen for the selected file.
3. Use the Arrow keys on the keyboard to move the cursor to the desired entry
field.
4. Type the numbers and press the Enter key on the keyboard to save each entry
and advance the cursor to the next entry field.
5. Repeat step 4 until all entries have been made.
6. If desired, press
[PRINT]
to obtain a printout of the entered values.
7. Press
[RETURN]
to save the entries and return to the
QC SET UP
screen.
NOTE:
When the entries are saved by pressing the
[RETURN]
soft key, the
software checks to see if any entries would result in the upper limit
being less than the lower limit. If this situation occurs, the limits are
automatically reversed. The bulletin line displays the message:
LIMITS WERE EXCHANGED TO MAKE UPPER > LOWER
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...