CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.4 Mixing Contact Sensor Check/Adjustment
3 - 97
3
7. If Step 5 or 6 fails, loosen screws on Sensor Bracket and adjust Contact Sensor.
Figure 3-50 [NON-CE MARK] Contact Sensor to Mixing Paddle Clearance
0.010 -0.015 in.
(0.015 Preferred)
(0.254-.381mm)
Sensor
Push Up
With Finger
Mixing
Paddle
SL-200E