28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURES
VP - 2 Keyboard Test
3 - 16
3
VP - 2 Keyboard Test
Purpose: Perform this procedure to verify Keyboard, Status LED, and Alarm LED operation.
1. Turn Sample Loader OFF.
2. Observe all LEDs while turning on power. All should flash (
).
3. Remove Cover to Manual Keyboard.
4. Set BAR (1/0) Switch to 0.
5. Verify PRE rotary switch (if used) is set to 0 (painted dot faces 0).
6. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
7. Press each Keyboard key. Verify BEEPING sounds and corresponding LED turns ON when key
is pressed.
Special Tools
DESCRIPTION
QTY
PN
None