![Abbott CELL-DYN 3000 Скачать руководство пользователя страница 494](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973494.webp)
CELL-DYN
3000 System Operator’s Manual
10-65
9140240E — May 1995
Chapter 10
Troubleshooting
Table 10.2
General Fault Conditions (Continued)
Auto-Sampler alarm condition <x,x> see DIAG/BL
NOTE: The characters in the brackets identify the condition.
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
The Sample Loader
1.
From the first
detected a hardware fault
and ceased operation.
, press [FAULT
REPORT] followed by
[PRINT] to obtain a
printout describing the
problem. Initialize the
Sample Loader by pressing
the INIT key on the Sample
Loader operation keyboard.
Samples may be processed
if the fault does not recur.
2.
Circuitry malfunction.
2.
Call for technical assistance.
__________________________________________________________
Auto-Sampler command negatively acknowledged/BL
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
The Sample Loader did
1.
Ensure that the Sample
not respond to an Analyzer
Loader power cord is
command.
connected to the Sample
Loader and that the cord is
connected to the power
outlet. Initialize the Sample
Loader by pressing the INIT
key on the Sample Loader
operation keyboard.
Check the connections of
the cable that connects the
Sample Loader to the
Analyzer. If necessary,
secure the connections.
Initialize the Sample Loader
by pressing the INIT key on
the Sample Loader
operation keyboard.
2.
Circuitry malfunction.
2.
Call for technical assistance.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......