Troubleshooting
Chapter 10
Table 10.2
General Fault Conditions (Continued)
External waste full/SB
WARNING: Potential Biohazard. Follow established
biosafety practices when performing maintenance, service or
troubleshooting procedures. Wear gloves, lab coat, and suitable
eye protection. Refer to
Chapter 8, Precautions, Limitations and
– – – – – Probable Cause – – – – –
– – – – Corrective Action – – – –
1.
Waste container full.
1.
Empty the waste container
and/or replace it. Press
[CLEAR FAULT] to resume
operation.
2.
Waste sensor connector
2.
Reconnect the waste sensor
is loose or disconnected.
connector and then press
[CLEAR FAULT].
3.
Shorted wire(s) or
3.
Inspect wires and
electrode(s) on the waste
electrodes and call
cap.
for technical assistance.
4.
Circuitry malfunction.
4.
Call for technical assistance.
__________________________________________________________
Blood in Line/BL
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
At the beginning of the
1.
Check the sample line to see
count cycle, the instrument
if it contains blood. If it
detects blood in the sample
does, clean or replace the
line that runs from the top
Sample Loader needle.
of the Sample Loader tower
to the shear valve.
2.
After four Blood in Line
2.
Check for a crimp in the
faults, the Sample Loader
sample aspiration tubing
halts.
between the Sample Loader
and the Analyzer.
NOTE: If necessary, samples can be processed in the open mode,
as it is not affected by this problem. To run samples in the open
mode, first initialize the system as directed in the
this chapter. When initialization is complete, the open mode is
automatically selected and samples can then be processed.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......