![Abbott CELL-DYN 3000 Скачать руководство пользователя страница 444](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973444.webp)
CELL-DYN
3000 System Operator’s Manual
10-15
9140240E — May 1995
Chapter 10
Troubleshooting
This key is useful for troubleshooting pressure problems. If the pump
does not turn ON when the key is pressed, the pump may be the cause of
the pressure problem.
When the [INHIBIT PUMPS] key is pressed, the key label changes to
[ENABLE PUMPS], operation of the pumps is inhibited (no vacuum or
pressure is produced) and the screen displays the appropriate message.
(See Figure 10.12.) Press the [ENABLE PUMPS] key to enable pump
operation.
NOTE: The pumps are automatically enabled and control of
them is returned to the instrument when the screen is exited.
This key is useful when performing maintenance or troubleshooting
procedures that require a vacuum or pressure line to be removed.
Figure 10.12:
Inhibit Pumps Screen
When the [VACUUM TEST] key is pressed, the system releases the
vacuum to atmosphere and then determines the amount of time required
for it to return to the correct level. The key labels disappear and the
incrementing time is displayed on the screen
the test is complete, the time stops incrementing and the key labels are
displayed.
INHIBIT
PUMPS
VACUUM
OFF
PRESSURE
OFF
ENABLE
PUMPS
VACUUM
TEST
PRESSURE
TEST
DIAG-
NOSTICS
DIAGNOSTICS MENU
Not Ready: See DIAG
Jan 18 1993
Operator ID
Sequence #
12:36
734
3143
Vacuum is inhibited
VACUUM
TEST
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......