5-96
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
5.
Run the sample in the Open mode.
6.
When the cycle is completed, press [CHANGE SAMPLER] to
select the Closed mode.
7.
Press the START key on the Sample Loader. Processing resumes
with the next sample to be run.
Sample Analysis with the Work List,
Closed Sampler Instruments
The Work List may be used in the Open and Closed modes on the Closed
Sampler version of the CELL-DYN 3000. On CS instruments, use of the
Work List is identical in both modes so they are discussed together.
Using the Work List with Bar Code ON
When the Bar Code ON option is selected, the Laboratory Specimen ID
option should always be selected in the Work List Set Up. The 4-Digit
Bar Code option should not be used.
When the Bar Code ON option is selected, the bar code labeled samples
may be run in any order after the Work List has been created. The
specimen ID number on the tube can be entered manually in the <NEXT
ID> entry field on the
in the Data Log by this number.
The Work List is accessed randomly as the samples are processed.
Sample identification is made by matching the Specimen ID number
entered on the
to the corresponding Work List
information. This information is used to identify the sample on the
NOTE: QC samples should not be entered in the Work List. To
run QC samples turn the Work List OFF and refer to the “
” section of this chapter for instructions
for running QC samples.
Sample Analysis Procedure
1.
Follow the instructions in the
to
configure the Work List.
NOTE: Be sure to select the Laboratory Specimen ID option on
the
2.
, press [WORK LIST ON] to
enable the Work List.
3.
Press [BAR CODE ON] to enable the bar code function.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......