TB82 TRANSMITTER
CALIBRATION
N
-
5
One-point calibration conducts an offset adjustment on the sensor
input. Use this type of calibration when the sensor is in the final
installed location.
Key
Display
Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select
YES
.
Select a calibration type.
Throughout this procedure the variable nnnn repre-
sents one of the following device types:
pH
ORP
PION
CONC
Select
1 POINT CAL
. Use this calibration method
when the sensor is in its final location.
Displays the current input value.
Enter the desired value.
The terminal displays the calculated efficiency in per-
centage.
C A LI-
B R AT E
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→
YES
E N T E R
→
nnnn CAL
TEMP. CAL
EDIT CAL
RESET CAL
E N T E R
nnnn CAL
→
1 POINT CAL
2 POINT CAL
E N T E R
nn.nn UNITS
PRESS ENTER
WHEN STABLE
E N T E R
ENTER NEW VALUE
_ nn.nn UNITS
E N T E R
CALCULATED
EFFICIENCY
nn.n %
PRESS ENTER
Содержание STT04
Страница 1: ...INSTRUCTION MANUAL Smart Transmitter Terminal STT04 PN25054 ...
Страница 179: ...TBN580 TRANSMITTER CALIBRATION K 12 Key Display Comments ENTER ID TAGNAME READY ...
Страница 253: ...AS800 PRESSURE TRANSMITTER CALIBRATION P 11 TEMPERATURE CALIBRATION is complete ENTER ID TAGNAME READY ...