TBN580 TRANSMITTER
CALIBRATION
K
-
7
Bench Calibration pH
The following procedure steps through the bench calibration of the pH
level.
Value from last calibration. If
NEW SENSOR
, effi-
ciency will be 98%. This value can be changed if
desired. A change in efficiency will affect the calibra-
tion span.
The current PH value is displayed. Enter the desired
PH value.
This value is recalculated during each calibration.
This value shows the offset change over time for a
particular sensor. Value resets to 0.00 pH if installing
a new sensor.
Key
Display
Comments
E N T E R
CHANGE
EFFICIENCY?
nn.nn%
HIT ENTER
E N T E R
CURRENT READING
nn.nn pH
ENTER DESIRED
nn.nn pH
E N T E R
CALCULATED
EFFICIENCY
nnn.nn%
HIT ENTER
E N T E R
CALCULATED
SENSOR OFFSET
nnn.nn pH
HIT ENTER
E N T E R
CUMULATIVE
OFFSET CHANGE
n.nn pH
HIT ENTER
E N T E R
ID TAGNAME
READY