TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
J - 8
Editing the Calibration Constants
These functions are for recording or modifying slope or offset calibra-
tion constants.
EDIT CONDUCTIVITY
This procedure steps through the modification of the conductivity.
Use the numeric keys to enter the process
temperature.
Key
Display
Comments
E N T E R
CHANGE TEMP FOR CALI-
BRATION
nn.nn °C
E N T E R
WORKING
CALIBRATION
COMPLETE
E N T E R
ID TAGNAME
READY
Key
Display
Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select
YES
.
C A LI-
B R AT E
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→
YES
Содержание STT04
Страница 1: ...INSTRUCTION MANUAL Smart Transmitter Terminal STT04 PN25054 ...
Страница 179: ...TBN580 TRANSMITTER CALIBRATION K 12 Key Display Comments ENTER ID TAGNAME READY ...
Страница 253: ...AS800 PRESSURE TRANSMITTER CALIBRATION P 11 TEMPERATURE CALIBRATION is complete ENTER ID TAGNAME READY ...