A ppENDIX C
Gas phase compensation static factors
The following tables explain provide the necessary gas phase compensation (GPC) static factors.
Dielectric constant of steam
Temperatrure
Pressure
°C
°F
1 bar
14.5 psi
2 bar
29 psi
5 bar
72.5 psi
10 bar
145 psi
20 bar
290 psi
50 bar
725 psi
100 bar
1450 psi
200 bar
2900 psi
100
212
1.005806
120
248
1.005227
1.010601
152
306
1.004476
1.009048
1.023424
180
356
1.003950
1.007964
1.020432
1.042934
212
414
1.003458
1.006960
1.017743
1.036765
1.079856
264
507
1.002840
1.005705
1.014456
1.029597
1.062307
1.192220
311
592
1.002418
1.004851
1.012252
1.024933
1.051729
1.147384
1.424747
366
691
1.002036
1.004082
1.010283
1.020834
1.042799
1.116952
1.282623
3.086361
Example with steam: resulting error in the gas phase
Gas phase
Temperatrure
Pressure
°C
°F
5 bar
72.5 psi
5 bar
72.5 psi
5 bar
72.5 psi
5 bar
72.5 psi
5 bar
72.5 psi
5 bar
72.5 psi
5 bar
72.5 psi
5 bar
72.5 psi
Steam
(water vapor)
100
212
0.26%
120
248
0.23%
0.50%
152
306
0.20%
0.42%
1.14%
180
356
0.17%
0.37%
0.99%
2.10%
212
414
0.15%
0.32%
0.86%
1.79%
3.9%
264
507
0.12%
0.26%
0.69%
1.44%
3.0%
9.2%
311
592
0.09%
0.22%
0.58%
1.21%
2.5%
7.1%
19.3%
366
691
0.07%
0.18%
0.49%
1.01%
2.1%
5.7%
13.2%
79%
Содержание LWT300 series
Страница 1: ... USER GUIDE LWT300 series Guided wave radar level transmitter Modbus ...
Страница 36: ...Page intentionally left blank ...
Страница 40: ...Page intentionally left blank ...
Страница 66: ...Page intentionally left blank ...
Страница 70: ...Page intentionally left blank ...
Страница 76: ...Page intentionally left blank ...
Страница 96: ...Page intentionally left blank ...
Страница 98: ...Page intentionally left blank ...
Страница 104: ...Page intentionally left blank ...
Страница 105: ...Page intentionally left blank ...
Страница 106: ...Page intentionally left blank ...
Страница 107: ......