Information INF09/027 Rev. D 

Endura AZ20 oxygen monitor

Pumped reference air unit

1 Introduction

This publication details the installation, connection and

maintenance of an ABB Pumped Reference Air Unit. This pump

supplies reference air to any of the following ABB oxygen probes:

AZ20, AZ25, AZ20/ZFG2 replacement, ZFG2 and ZGP2

4 pump models are available – see Table 1.1:

The internal air filter is a replaceable item (ABB part

number 0217463).

2 Safety

Information in this document is intended only to assist our

customers in the efficient operation of our equipment. Use of this

manual for any other purpose is specifically prohibited and its

contents are not to be reproduced in full or part without prior

approval of the Technical Publications Department.

2.1 Safety precautions

Please read the entire document before unpacking, setting up, or

operating this pump. 

Pay particular attention to all warning and caution statements.

Failure to do so could result in serious injury to the operator or

damage to the equipment. 

To ensure the protection provided by this equipment is not

impaired, do not use or install this equipment in any manner other

than that which is specified in this manual.

2.2 Health & Safety

Note.

The pump cannot be used with multiple probes

simultaneously.

Description

ABB Part Number

1

/

4

 BSP (Metric) 230 V AC 50 / 60 Hz

AZ200 770

1

/

4

 BSP (Metric) 115 V AC 50 / 60 Hz

AZ200 771

1

/

4

 NPT (Imperial) 230 V AC 50 / 60 Hz

AZ200 772

1

/

4

 NPT (Imperial) 115 V AC 50 / 60 Hz

AZ200 773

Table 1.1 ABB Pumped Reference Air Unit Part Numbers

To ensure that our products are safe and without risk to health,

the following points must be noted:

The relevant sections of these instructions must be read

carefully before proceeding.

Warning labels on containers and packages must be

observed.

Installation, operation, maintenance and servicing must

only be carried out by suitably trained personnel and in

accordance with the information given.

Normal safety precautions must be taken to avoid the

possibility of an accident occurring when operating in

conditions of high pressure and / or temperature.

Safety advice concerning the use of the equipment described

in this manual or any relevant Material Safety Data Sheets

(where applicable) may be obtained from the Company address

on the back cover, together with servicing and spares

information.

Содержание AZ200 770

Страница 1: ...uipment is not impaired do not use or install this equipment in any manner other than that which is specified in this manual 2 2 Health Safety Note The pump cannot be used with multiple probes simultaneously Description ABB Part Number 1 4 BSP Metric 230 V AC 50 60 Hz AZ200 770 1 4 BSP Metric 115 V AC 50 60 Hz AZ200 771 1 4 NPT Imperial 230 V AC 50 60 Hz AZ200 772 1 4 NPT Imperial 115 V AC 50 60 H...

Страница 2: ...he enclosure or remove the barrier This symbol indicates that the marked item can be hot and should not be touched without care This symbol indicates the presence of devices sensitive to electrostatic discharge and indicates that care must be taken to prevent damage to them The European Union RoHS Directive and subsequent regulations introduced in member states and other countries limits the use o...

Страница 3: ...oblems with installation starting or using the pump please contact the company that sold it to you If this is not possible or if the results of this approach are not satisfactory please contact the manufacturer s Customer Service 2 9 Potential safety hazards The following potential safety hazards are associated with operating the system Electrical line voltage Warning The installation of the pump ...

Страница 4: ... with relevant national and local standards Use cable appropriate for the load currents 3 core cable rated 5A and 90 C 194 F minimum that conform to either IEC 60227 or IEC 60245 The terminals accept cables from 0 8 to 2 5 mm2 18 to 14 AWG The pump conforms to Installation Category II of IEC 61010 After installation there must be no access to live parts for example terminals If the pump is used in...

Страница 5: ...osen cable gland C to prepare for the incoming power supply cable 3 5 Electrical connections 3 5 1 Preparation Referring to Fig 3 4 1 Prepare the incoming power supply cable as follows a cut the outer sheath back to leave 100 mm 4 0 in sleeved wires A b cut the sleeved wires back to leave 5 mm 0 2 in tails B 2 Pass the cable through cable gland C Fig 3 3 Accessing the Terminal Block B C A Pump Ide...

Страница 6: ...ction 6 Refit the pump cover by reversing the procedure in Section 3 4 3 6 Reference air connections Referring to Fig 3 5 1 It is important that air to the pump inlet is clean and dry If necessary connect a pipe to the pump suction inlet A and ensure the other end of the pipe is located in a clean dry area 2 Connect the pumped reference air outlet B directly to the reference air inlet on the probe...

Страница 7: ...ring to Fig 4 1 2 Disconnect the top tube A from air filter B 3 Disconnect the lower tube C from air filter B 4 Release clip D and carefully remove air filter B by pulling it out of the clip 5 Discard air filter B 6 Fit a new filter by reversing steps 1 to 4 Note Isolate the pump from power and air supplies before replacing the air filter Fig 4 1 Replacing the Air Filter A C B D ...

Страница 8: ...or modify the contents of this document without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or ut...

Отзывы: