Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
5 PROGRAMMING
46
IM/AX4CO Issue 11
5.8.2 Configure Power Failure Recovery Mode
Power Failure Recovery Mode
When power to the analyzer is restored,
Control Mode
(Section 2.3) is set automatically
according to the Power Failure Recovery Mode selected in this frame.
Select the required mode:
Auto
Manual
Last
-
-
-
Control Mode
is set to
Auto
irrespective of its setting prior to the
power failure.
Control Mode
is set to
Manual
irrespective of its setting prior to the
power failure.
Control Output
(Section 2.3) is set to the level set in
the
Default Output
frame below.
Control Mode
and
Control Output
are set to the same state as that
set prior to the power failure.
Default Output
Set the default output required after Power Failure Recovery, between 0 and 100% in
0.1% increments.
Note.
A setting of 0% represents no output.
Return to main menu.
Power Recovery
----
Power Rec. Mode
Default Output
50.0
----
%
Last
Manual
Auto
CONFIG. CONTROL
Power Recovery
CONFIG. SECURITY