pH/Redox/ORP
This section describes how to calibrate the sensor and involves measuring the
sensor’s sensitivity to pH and temperature by exposing the sensor to samples of
known pH/Temperature values
Notes
.
• Access to the
Calibrate
menu is via
Calibrate
and
Advanced
levels only.
• During calibration, current outputs and alarms are set to hold automatically
if
Hold Outputs
is enabled – see below.
Menu
Comment
Default
Sensor Cal
See
pH/Redox/ORP calibration
There are four possible calibration modes:
• 1-point manual calibration (adjusts the calibration check value)
• 2-point manual calibration (adjusts the check and slope values)
• 1-Point automatic calibration (adjusts the calibration check value)
• 2-Point automatic calibration (adjusts the check and slope values)
Note .
Automatic calibrations are not available for Redox/ORP measurements
Temperature Cal
See
Temperature calibration procedure
, page 69
Edit Cal
pH Slope
Note:
pH sensors only.
Edit the slope value:
• Valid slope values range from 40 to 150 %
100%
pH Offset
Note:
pH sensors only.
Edit the offset value:
• Valid offset values are: 0.00 to 14.00pH
7.00pH
mV Slope
Note:
Redox/ORP sensors only.
Edit the slope value:
• Valid slope values range from 40 to 150 %
100%
mV Offset
Note:
Redox/ORP sensors only.
Edit the offset value:
• Valid offset values are: ±1000mV
0mV
Temp . Slope
Edit the temperature slope value:
• Valid slope values range from 40 to 160 %
100%
Temp . Offset
Edit the temperature offset value:
• Valid offset values are ±40 °C
0°C
Sample Collection
Note:
pH sensors only.
See
In process calibrations
Collection Complete
Note:
pH sensors only
See
In process calibrations
Restore Cal Defaults
Resets slope and offset values to default values.
Calibrate
Exit
Menu
Select