2- W I R E CO N D U C T I V I T Y, P H/O R P, P I O N T R A N SM I T T ER | O I/AW T 2 10 - EN R E V. C
11 Configuration
Easy Setup
pH
Used to setup standard parameters quickly.
Note.
Easy Setup
menus are sensor-specific – the menus in this section are applicable to
transmitters connected to pH sensor types.
Menu
Comment
Default
Language
Selects the language to be displayed on screen:
English/Deutsch/Français/Español/Italiano
Note. The language options are displayed with localized spelling.
Measurement Type
Selects a measurement type compatible with the connected sensor:
pH/ORP/Ion Conc/pION
pH Sensor Type
Select a measurement type compatible with the connected sensor:
Glass/Antimony/Custom
Isopotential Pt .
Isopotential point.
Enabled only if
Measurement Type
=
pH
and
pH Sensor Type
=
Custom
.
Asymmetric Pot .
Asymmetric potential.
Enabled only if
Measurement Type = pH
and
pH Sensor Type = Custom.
PV Unit
Sets the process value units.
Enabled only if
Measurement Type = Ion Conc.
%/ppm/ppb/µg/l/mg/l
Valence
The ion valence determines the millivolt change per decade of concentration.
Enabled only if
Measurement Type = Ion Conc.
–3/–2/–1/1/2/3
Magnitude
The number of magnitudes (ranging from 1 to 3) that define the ion concentration output.
Enabled only if
Measurement Type
=
Ion Conc. 1/2/3
End Magnitude
The ion concentration state functions by associating an end millivolt
value to an end magnitude value.
Enabled only if
Measurement Type = Ion Conc.
10/100/1000
End mV
The end millivolt value associated with the end magnitude value.
Enabled only if
Measurement Type = Ion Conc
.
Temp . Unit
Temperature unit displayed and applied to all temperature parameters/values.
Not enabled if
Temperature Compensation Type = Manual
.
°F/°C
Temp . Comp . Type
Temperature compensation type – see page 51.
Manual/Auto/Auto Solution
Note. When
Auto
is selected and a temperature sensor is recognized, the temperature compensation
calculation is based on the Nernst equation detailed on page 51. When
Auto Solution
is selected and a
temperature sensor is provided with the probe, the calculation is based on the Nernst solution coefficient
equation detailed in detailed on page 51. When
Manual
is enabled/selected, the value set in the
Manual
Temp.
parameter is used for the referenced temperature compensation.
Auto
(if valid
temperature
sensor
recognized)
Manual Temp .
Temperature value used for compensation in
Temp Comp. Type/Manual
compensation mode.
Enabled only if
Temperature Compensation Type = Manual.
pH Soln . Coeff .
Enabled only if
Measurement Type = pH
and
Temp. Comp. Type = Auto Solution
.
mV Soln . Coeff .
Enabled only if
Measurement Type = pH
and
Temp. Comp. Type = Auto.
Operator Page 1
Selects range of values displayed on
Operator Page 1
– see page 47 for examples of character formats/sizes.
PV/PV + Temperature/PV + Temp + PV%/PV + Temp + O/P
Easy Setup
Exit
Menu
Select