2-electrode conductivity
Used to setup standard parameters quickly.
Note.
Easy Setup
menus are sensor-specific – the menus in this section are applicable to
transmitters connected to 2-electrode conductivity sensor types.
Menu
Comment
Default
Language
Selects the language to be displayed on screen:
English/Deutsch/Français/Español/Italiano
Note. The language options are displayed with localized spelling.
Measurement Type
Selects a measurement type compatible with the connected sensor:
Conductivity/Concentration
Cell Constant
Used to adjust the slope of the calculation graph.
The design of the conductivity sensor determines the sensor constant (usually k = 1).
Conc . Unit
Sets the process value units.
Enabled only if
Measurement Type = Concentration
.
%/ppm/ppb/µg/l/mg/l
Conc . Curve Name
Enabled only if
Measurement Type = Concentration
.
Temp . Unit
Temperature unit displayed and applied to all temperature parameters/values.
Not enabled if
Temperature Compensation Type = Manual
.
°F/°C
Operator Page 1
Selects range of values displayed on
Operator Page 1
– see page 47 for examples of character formats/sizes.
PV/PV + Temperature/PV + Temp + PV%/PV + Temp + O/P
Easy Setup
Exit
Menu
Select