![ABB AO2000-LS25 Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/abb/ao2000-ls25/ao2000-ls25_manual_4322023.webp)
AO2000-LS25
LASER ANALYZERS | OI/AO2000-LS25-EN REV. D
23
CSA certification, gas analyzer
CSA Class I Division 2 Groups A, B, C, D, Temperature class T4,
electrical equipment
Applicable requirements:
• CAN/CSA C22.2 No. 0-M91 (R2001): General Requirements
– Canadian Electrical Code, Part II
• CSA standard C22.2 No. 142-M1987:
Process Control Equipment
• CSA standard C22.2 No. 213-M1987:
Non-lncendive Electrical Equipment for Use in Class I,
Division 2 Hazardous Locations
• UL standard No. 916-2007: Energy Management
Equipment
• ANSI/ISA-12.12.01-2010:
Non-lncendive – Electrical Equipment for Use in Class I
and II, Division 2 and Class III, Divisions 1 and 2 Hazardous
(Classified) Locations
Certificate no.:
1105720
CSA certification, power supply unit
The power supply unit has its own CSA approval, which is
registered under UL File e213613 (Class I, Division 2, group A, B, C
& D, temp. Class T4).
Ambient conditions
Operating temperature
−
20 °C to +55 °C
Storage temperature
−
20 °C to +55 °C
IP rating
Designed for outdoor applications (all modules IP 66-
certified).
IP rating is not part of the CSA certification
Degree of contamination
2
Installation category (overvoltage category)
II
Installation height
2000 m
Air humidity
0 to 95 %
Electrical data
Power supply
Power supply unit (input)
85 to 264 V AC, 50 / 60 Hz
Power supply unit (output)
24 V DC, maximum 5 A
Transmitter unit (input)
18 to 32 V DC, maximum 20 W
Interfaces
Service interface
• RS232 according to EIA/TIA-232 and V.28/V.24
• Level: max. ±15 V.
• Current (short-circuit): max. ±60 mA
Ethernet interface
• Protocol 10/100Base-T
• Max. voltage: ±2.5V
Safety instructions
DANGER
Explosion hazard
Failure to follow safety instructions may cause ignition of
hazardous atmosphere resulting in serious personal injury
and/or property damage.
WARNING
Loss of Ex-approval!
Loss of Ex approval due to replacement of components in
devices for use in potentially explosive atmospheres.
• Devices for use in potentially explosive atmospheres may
be serviced and repaired by qualified ABB personnel only.
• For measuring devices for potentially explosive
atmospheres, observe the relevant operator guidelines.