2 OPERATION
IM/AX4PH
Issue 9
7
2.3.2 Dual Input pH
Measured pH
Sensor A.
Sensor B.
Measured Temperature
Sensor A.
Sensor B.
Measured Millivolts
Sensor A.
Sensor B.
% Slope and pH Check Value - Sensor A
% slope value.
A value between the programmed minimum % slope value (see
Set Min Slope
section 5.3, page 30) and 105% is displayed. If the value is outside these limits, check
the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7pH being the
optimum value for glass electrodes and 0pH for Antimony electrodes.
% Slope and pH Check Value - Sensor B
% slope value.
See Sensor A above.
pH check value (zero point).
See Section 3.1.
See Section 4.1.