4690 Series
Turbidity systems
8 Configuration and Electrical Calibration
28
IM/4690–EN Rev. A
Continued from previous page
Cleaning Interval
Note.
Displayed only if
Sensor Type
is set to
7998-001, 7998-002, 7998-011
or
7998-012
Using the
and
keys, select the cleaning interval required. Values available are: 15 mins,
30 mins, 45 mins, 1 hr, then 1 hour increments to a maximum of 24 hours.
Test Cleaner
Note.
Displayed only if
Sensor Type
is set to
7998-001, 7998-002, 7998-011
or
7998-012
Select
YES
to activate the wiper unit diagnostics function and enable visual indication of wiper unit
function.
Select
No
to deactivate the wiper unit diagnostics function.
Bubble Rejection
Use in applications where degassing of the sample takes place.
When set to
YES
, the analyzer ignores short duration spikes in the turbidity reading caused by the
formation of small bubbles in the flowcell due to degassing.
Bubble Period
Note.
Displayed only if
Bubble Rej.
is set to
YES
Using the
and
keys, adjust the value to between 2 and 60 seconds until short duration
spikes are no longer seen on the display.
Note.
A value of 20 seconds is normally sufficient for most applications.
Signal Filter
Use in applications where the turbidity readings are noisy.
Using the
and
keys, adjust the value to between 4 and 180 seconds.
Note.
If
Bubble Rej.
is set to
YES
, set a value of 4 seconds for optimum performance.
Advance to
Set Up Outputs
page – see Section 8.4, page 29.
YES
Test Cleaner NO
YES
1 3
Clean Int. (Hr)
SET UP OUTPUTS
- - - - -
Bubble Rej. NO
Bubble Period
2 0
Signal Filter
2 0
Return to
top of page