Quality Control and Calibration
5-11
5.2.3
Calibration Procedure
The calibration process requires at least three sample runs using the HM5 Normal control material.
1.
From the Home screen, select
Maintenance > Calibration
.
2.
Select
Measure
.
3.
Verify that
Prediluted
is set to
No
and that
Cal Sampling Depth
is set to
–2 mm
.
4.
Select
Accept
.
Note:
Allow the normal control to reach room temperature before beginning.
This takes about 15 minutes.
Once opened, the control has a shelf life of 14 days. Do not calibrate
with an expired control or a tube that has been opened more than 14
days.