4.5 Running a Sample
Use this general procedure to analyze samples with the VetScan HM5.
1.
Prepare a well-mixed, potassium EDTA-preserved sample — see
.
2.
Select the appropriate sample tube adapter for the tube size being used:
#1:
for 2-3 ml tubes and the Abaxis Control tube.
#2:
for 1.3 ml tubes.
#3:
for 2 ml glass vials.
#6:
for microtainer tubes of 0.6 ml or less
3.
Place the adapter into the slot in the sample door.
4.
From the Home screen, select
Measure > Run
.
5.
Select the patient species under
Type
.
6.
Enter patient information: select the needed fields, enter the infor-
mation, then select
Enter
.
7.
Make sure the sample tube is at least half full. If the tube is not
adequately filled (see
“Collecting and Handling Samples” on
), results can be affected.
Good sample acquisition and
sample handling are essential to prevent issues.
Note:
If combining with a VS2 or sending results to
a PMS, be sure to enter the matching patient
number into the Patient ID.
Содержание VetScan HM5
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 4: ...1 2 ...
Страница 6: ...1 4 ...
Страница 26: ...2 10 Installing the VetScan HM5 ...
Страница 38: ...3 12 Settings ...
Страница 50: ...4 12 Running a Sample ...
Страница 64: ...5 14 Quality Control and Calibration ...
Страница 80: ...6 16 Maintenance Service ...
Страница 88: ...8 4 Shutdown ...
Страница 112: ...9 24 Troubleshooting ...
Страница 126: ...A 10 Introduction to Veterinary Hematology ...
Страница 136: ...B 10 Operating Principles ...
Страница 154: ...D 16 Veterinary Case Studies ...
Страница 163: ......