2-6
Installing the VetScan HM5
19.
Select a
Pack Setup
:
If the reagent pack will be
on the
same level
as the HM5 (preferred) and
you have the QCA (Quick Change
Apparatus), select
Quick Change
(QCA)
in
Reagent Replace.
If the reagent pack will be
on a lower level
than the
HM5 and/or the Bottle Caps tubing set is being
used, select
Bottle Caps
. Make sure the top of the
reagent pack is no more than 18 in (45.7 cm) lower
than the bottom of the analyzer.
20.
Select
Schedule Settings > Reset > Accept
to restart maintenance timers from today.
21.
Pour ~200 ml of distilled water into the Abaxis flask.
22.
Submerge the free ends of the reagent tubes into the
water.
23.
Select
Maintenance > Reagent Status
, and then
Prime All
.
24.
Run
Prime All
twice more (three times in all) until all
water is gone from the flask and air is pumped into
the tubing.
25.
While the analyzer is priming, open the reagent pack.
Remove the bottle caps from each of the five bottles. Save these lids for later disposal of the
pack.
Содержание VetScan HM5
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 4: ...1 2 ...
Страница 6: ...1 4 ...
Страница 26: ...2 10 Installing the VetScan HM5 ...
Страница 38: ...3 12 Settings ...
Страница 50: ...4 12 Running a Sample ...
Страница 64: ...5 14 Quality Control and Calibration ...
Страница 80: ...6 16 Maintenance Service ...
Страница 88: ...8 4 Shutdown ...
Страница 112: ...9 24 Troubleshooting ...
Страница 126: ...A 10 Introduction to Veterinary Hematology ...
Страница 136: ...B 10 Operating Principles ...
Страница 154: ...D 16 Veterinary Case Studies ...
Страница 163: ......