Operating Instructions — Analyze and Process Quantitative Data
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
97 of 138
ii. In the
Internal Standard
column, from the list, select the internal standard to
be associated with each analyte.
iii. In the
Q1/Q3
column
,
select 609.400/195.039.
iv. If required, add one or more of the other compounds for a more complex
analysis.
7. Click the
Integration
tab.
In general, the preset integration parameters are suitable for most peaks.
8. If the integration is not suitable, then change the algorithm.
9. Click the
Show or Hide Parameters
icon to show the additional integration
algorithms.
10. Click the
Calibration
tab. The preset parameters are suitable for these samples.
11. Save the quantitation method.
The new method can be used when a batch is created in the
Batch Editor
or when
the
Quantitation Wizard
is used to generate a Results Table.
Create a Results Table using the Quantitation Wizard
It is recommended that the user validate any queries that are used to analyze data in a Results
Table.
1. In the navigation bar, under
Quantitate
, double-click
Quantitation Wizard
.
The
Create Quantitation Set - Select Samples
page opens.
2. In the
Available Data Files
list, double-click the
Triple Quad
folder.
3. Select
Mix_Batch_2.wiff
.
4. Click
Add All Files
.
5. Click
Next
.
The
Create Quantitation Set - Select Settings & Query
page opens.
6. In the
Default Query
section, click
Select Existing: Query
.
7. In the
Query
list, select
Accuracy 15%
.
Note:
If the
Compound ID
field was populated for the samples and
internal standards in the acquisition method, then in the Analytes table, the
Name
field and
Q1/Q3
field are populated.
Note:
The quantitation method can only be used in the current project
unless it is copied to another project. To do this, click
Tools > Project >
Copy Data
. A new project must be created and selected to be available for
use.
Prerequisite Procedures
• Select the
Example
project. Refer to
Switch Between Projects and Subprojects on page 47
.
• Use the
Analyst Classic
algorithm. Refer to
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...